EZ1005 Chloride Analyser

EZ1005 Chloride Analyser
Nr. produs: EZ1005.XXXXXXXX
Preț unitar Contact

Conţinutul cutiei

EZ1005 Chloride Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 2 x empty 2.5L Reagent Container with Fittings (for Acid & Cleaning Solution) and 1 x empty 2.5L Glass Reagent Container with Fittings (for AgNO₃ Solution)

Specificaţii

Alarma: 1x malfunctioning, 4x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Alimentare electrică: 100 - 240 VAC, 50/60 Hz
Calibrare: Automatic, 2-point; frequency freely programmable
Calitate proba: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Certificări: CE compliant / ETL certified
Clasa de protectie: Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Conţinutul cutiei: EZ1005 Chloride Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 2 x empty 2.5L Reagent Container with Fittings (for Acid & Cleaning Solution) and 1 x empty 2.5L Glass Reagent Container with Fittings (for AgNO₃ Solution)
Curatare automata: Da
Cuve: 10 min (dilution + 5 min)
Debitul probei: 100 - 300 mL/min
Demineralised water: For rinsing / dilution
Digital outputs: Optional: Modbus (TCP/IP, RS485)
Dimensiuni (Înălţime x Lăţime x Adâncime): 690 mm x 465 mm x 330 mm
Domeniu de măsurare: 1 ‐ 10 mg/L
  1 - 10 mg/L Cl-

Optional:
4 - 40 mg/L Cl- (with internal dilution)
8 - 80 mg/L Cl- (with internal dilution)
100 - 1000 mg/L Cl- (with internal dilution)
Earth connection: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Esantion: presiune: By external overflow vessel
Garanţie: 24 luni
Greutate: 25 kg
Iesiri analogice: Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferente: Substances in amounts normally found in potable water will likely not interfere. Bromide, iodidie and cyanide register as equivalent chloride concentrations. Sulfide, thiosulfate and sulfite interfere but can be removed by treatment with hydrogen peroxide. Orthophosphate > 25 mg/L interferes by precipitating as silver phosphate. Iron > 10 mg/L interferes. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Limita de detecţie: 1 mg/L
Material: Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Metoda de măsurare: Colorimetric measurement at 546 nm using diphenylcarbazide method, conform with standard method APHA 3500-Cr (B)
Model: EZ1005
Nr. flux probe : 1, 2, 4, or 8
Nume metoda: Colorimetric
Parametru: Chloride
Precizie: Better than 2% full scale range for standard test solutions
Putere: 100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Reagent Requirements: Keep between 10 - 30 °C
Scurgere: Atmospheric pressure, vented, min. Ø 32 mm
Temparatura ambiantă: 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura probă: 10 - 30 °C
Validation: Automatic; frequency freely programmable