Alarma:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Alimentare electrică:
100 - 240 VAC, 50/60 Hz
Calibrare:
Automatic, 2-point; frequency freely programmable
Calitate proba:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Certificări:
CE compliant / ETL certified
Clasa de protectie:
Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Conţinutul cutiei:
EZ1010 Copper Analyser, Instruction Manual, 1 x Double Bit Door key, 1 x Mounting Brackets and 2 x empty 5L Reagent Containers with Fittings (for Colour & Acid Solution)
Curatare automata:
Da
Cuve:
10 min (dilution + 5 min)
Debitul probei:
100 - 300 mL/min
Demineralised water:
For rinsing / dilution
Digital outputs:
Optional: Modbus (TCP/IP, RS485)
Dimensiuni (Înălţime x Lăţime x Adâncime):
690 mm x 465 mm x 330 mm
Domeniu de măsurare:
0.03 - 3 mg/L Cu(II)
0.03 - 3 mg/L Cu(II)
Optional:
0.003 - 0.3 mg/L
0.02 - 0.75 mg/L
0.02 - 1.5 mg/L
0.2 - 12 mg/L (with internal dilution)
0.5 - 24 mg/L (with internal dilution)
3 - 300 mg/L (with internal dilution)
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Esantion: presiune:
By external overflow vessel
Garanţie:
24 luni
Greutate:
25 kg
Iesiri analogice:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferente:
Acidity, metal ions like Aluminium (III) > 10 mg/L, Cyanide, Hardness, Iron (III) > 10 mg/L, Nickel (II) and Silver (II). Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Limita de detecţie:
≤ 3 µg/L
Material:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Metoda de măsurare:
Colorimetric measurement at 546 nm using bicinchoninate method; conform respectively with USEPA Method 8506 and standard method APHA 3500-Cu (C)
Model:
EZ1010
Nr. flux probe :
1, 2, 4, or 8
Parametru:
Copper, dissolved
Precizie:
Better than 2% full scale range for standard test solutions
Putere:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Reagent Requirements:
Keep between 10 - 30 °C
Scurgere:
Atmospheric pressure, vented, min. Ø 32 mm
Temparatura ambiantă:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura probă:
10 - 30 °C
Validation:
Automatic; frequency freely programmable