Calibrare:
Automatic; 2-point; offset or slope; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced
Calitate proba:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Certificări:
CE, ETL certified to UL and CSA safety standards, UKCA
Clasa de protectie:
IP44
Conţinutul cutiei:
EZ1017sc Calcium Hardness Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 3 x empty 2.5L Reagent Containers with Fittings (for Colour, Acid, and EDTA Solution) and 1 x empty 5L Reagent Container with Fittings (for Buffer)
Curatare automata:
Yes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
Cuve:
No Dilution
Default: 10 min.
Continuous: 8 min.
Internal Dilution:
Default: 15 min.
Continuous: 10 min.
Debitul probei:
100 - 300 mL/min
Demineralised water:
Internal dilution option: Rinse water & Dilution water
Digital outputs:
Profibus DP or Modbus RTU
Dimensiuni (Înălţime x Lăţime x Adâncime):
688 mm x 460 mm x 340 mm
Domeniu de măsurare:
No Dilution
0.005 - 0.25 mg/L CaCO₃
0.005 - 0.5 mg/L CaCO₃
0.025 - 1 mg/L CaCO₃
Interal Dilution
0.125 - 5 mg/L CaCO₃
0.25 - 10 mg/L CaCO₃
0.625 - 25 mg/L CaCO₃
1.25 - 50 mg/L CaCO₃
1.875 - 75 mg/L CaCO₃
2.5 - 100 mg/L CaCO₃
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Esantion: presiune:
By external overflow vessel (open to atmospheric pressure)
Garanţie:
24 luni
Greutate:
Max. 35 kg
Ieşire:
Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP
Iesiri analogice:
Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard: 4, optional: 8
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar
Interferente:
Some metal ions interfere by causing fading or indistinct end points or by stoichiometric consumption of EDTA. Suspended or colloidal organic matter can interfere with the end point. Large quantities of color and turbidity interfere. Fats, oil, proteins, surfactants and tar interfere.
Limita de detecţie:
No Dilution
0.005 - 0.25 mg/L CaCO₃: 0.005 mg/L
0.005 - 0.5 mg/L CaCO₃: 0.005 mg/L
0.025 - 1 mg/L CaCO₃: 0.025 mg/L
Interal Dilution
0.125 - 5 mg/L CaCO₃: 0.125 mg/L
0.25 - 10 mg/L CaCO₃: 0.25 mg/L
0.625 - 25 mg/L CaCO₃: 0.625 mg/L
1.25 - 50 mg/L CaCO₃: 1.25 mg/L
1.875 - 75 mg/L CaCO₃: 1.875 mg/L
2.5 - 100 mg/L CaCO₃: 2.5 mg/L
Material:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Metoda de măsurare:
Hydroxynaphthol blue colorimetric method
Nr. flux probe :
Standard: 1
Optional accessory for 2, 4, or 8 sample streams
Parametru:
Hardness, Calcium
Precizie:
Better than 3% of full scale at:
0.005 - 0.25 mg/L CaCO₃
0.005 - 0.5 mg/L CaCO₃; and better than 2% of full scale for standard test solutions.
Putere:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Reagent Requirements:
Keep between 10 - 30 °C
Relee:
5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150)
Scurgere:
Atmospheric pressure, vented, min. Ø 32 mm
Temparatura ambiantă:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura probă:
10 - 30 °C
Validation:
Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly