Alarma:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Alimentare electrică:
100 - 240 VAC, 50/60 Hz
Calibrare:
Automatic, 2-point; frequency freely programmable
Calitate proba:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Certificări:
CE compliant / ETL certified
Clasa de protectie:
Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Conţinutul cutiei:
EZ1022 Hydrogn Peroxide Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and 2 x empty 2.5L Glass Containers with Fittings (for Colour and Buffer Solution)
Curatare automata:
Da
Cuve:
10 min (dilution + 5 min.)
Debitul probei:
100 - 300 mL/min
Demineralised water:
For rinsing / dilution
Digital outputs:
Optional: Modbus (TCP/IP, RS485)
Dimensiuni (Înălţime x Lăţime x Adâncime):
690 mm x 465 mm x 330 mm
Domeniu de măsurare:
5 - 500 µg/L H₂O₂
5 - 500 µg/L H₂O₂
Optional:
2 - 125 µg/L
2 - 250 µg/L
0.04 - 2.0 mg/L (with internal dilution)
0.08 - 4.0 mg/L (with internal dilution)
0.5 - 50 mg/L (with internal dilution)
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Esantion: presiune:
By external overflow vessel
Garanţie:
24 luni
Greutate:
25 kg
Iesiri analogice:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferente:
Oxidants like Bromine, Bromamine, Chlorine, Copper Cu(II) > 8 mg/L, Chromate, Iodine, Iodoamines, Iron Fe(III) > 20 mg/L, Nitrite, Ozone and Manganese may oxidize the DPD colour solution. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Limita de detecţie:
≤ 2 µg/L
Material:
Hinged part: Thermoform ABS, door: PMMA
Wall section: galvanised steel, powder coated
Metoda de măsurare:
Colorimetric measurement at 510 nm using DPD
Model:
EZ1022
Nr. flux probe :
1, 2, 4, or 8
Parametru:
Hydrogen Peroxide
Precizie:
Better than 2% full scale range for standard test solutions
Putere:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Reagent Requirements:
Keep refrigerated.
Scurgere:
Atmospheric pressure, vented, min. Ø 32 mm
Temparatura ambiantă:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura probă:
10 - 30 °C
Validation:
Automatic; frequency freely programmable