Alarma:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Alimentare electrică:
100 - 240 VAC, 50/60 Hz
Calibrare:
Automatic, 2-point; frequency freely programmable
Calitate proba:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Certificări:
CE compliant / ETL certified
Clasa de protectie:
Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Conţinutul cutiei:
EZ1032 Phosphate Analyser, Instruction Manual, 1 x Double Bit Door key, 1 x Mounting Brackets, 1 x empty 5 Reagent Container with Fittings (for Colour Solution), and 1 x empty 2.5L Glass Reagent Container with Fittings (for Reductor Solution)
Curatare automata:
Da
Cuve:
10 min (dilution + 5 min)
Debitul probei:
100 - 300 mL/min
Demineralised water:
For rinsing / dilution
Digital outputs:
Optional: Modbus (TCP/IP, RS485)
Dimensiuni (Înălţime x Lăţime x Adâncime):
690 mm x 465 mm x 330 mm
Domeniu de măsurare:
0.005 - 1 mg/L PO₄-P
0.005 - 1 mg/L PO₄-P
Optional:
0.001 - 0.1 mg/L
0.0025 - 0.25 mg/L
0.0025 - 0.5 mg/L
0.04 - 4 mg/L (with internal dilution)
0.08 - 8 mg/L (with internal dilution)
0.5 - 100 mg/L (with internal dilution)
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Esantion: presiune:
By external overflow vessel
Garanţie:
24 luni
Greutate:
25 kg
Iesiri analogice:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferente:
Arsenic (V), Chromium (VI), Copper (II) > 10 mg/L, Iron (III) > 10 mg/L, Sulphide > 2 mg/L, Vanadium, Silica > 60 mg/L. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Limita de detecţie:
≤ 1 µg/L
Material:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Metoda de măsurare:
Molybdate blue method (630 nm), conform with APHA 4500-P (C) and ( E)
Model:
EZ1032
Nr. flux probe :
1, 2, 4, or 8
Parametru:
Orthophosphate (PO₄-P)
Precizie:
1% full scale range for standard test
Putere:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Reagent Requirements:
Keep between 10 - 30 °C
Scurgere:
Atmospheric pressure, vented, min. Ø 32 mm
Temparatura ambiantă:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura probă:
5 - 30 °C
Validation:
Automatic; frequency freely programmable