Alarma:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Calibrare:
Automatic, 2-point; frequency freely programmable
Calitate proba:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Carcasă impermeabilă (IP) :
IP44
Certificări:
CE compliant / ETL certified
Clasa de protectie:
IP44
Conţinutul cutiei:
EZ2727sc Nickel Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and Empty Containers for Reagents with Fittings
Cooling water:
Flow rate approx. 5 L/h; temperature max. 30 °C; pressure max. 0.5 bar
Curatare automata:
Da
Cuve:
20 minutes (dilution + 5 min.)
Debitul probei:
100 - 300 mL/min
Demineralised water:
When needed for rinsing / dilution
Digital outputs:
None
Optional: Modbus (TCP/IP, RTU), Profinet, Profibus DP, Ethernet IP
Dimensiuni (Înălţime x Lăţime x Adâncime):
688 mm x 460 mm x 340 mm
Domeniu de măsurare:
0.01 - 0.25 mg/L
0.02 - 0.5 mg/L
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Esantion: presiune:
By external overflow vessel
Garanţie:
24 luni
Greutate:
40 kg
Iesiri analogice:
Active 4 - 20 mA max. 500 Ohm load, standard 4, max. 8 (option)
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferente:
Metal ions like Aluminium Al(III), Bismuth Bi(II), Cadmium Cd(II), Chromium Cr(III), Cobalt Co(II), Copper Cu(II), Iron Fe(II), Iron Fe(III), Lead Pb(II), Manganese Mn(II), Magnesium Mg(II), Mercury Hg(II), Palladium Pd(II), Platinum Pt(II), Silver Ag(I), Tin Sn(II), Zinc Zn(II). Large amounts of colour and turbidity interferes. Fats, oil, proteins, surfactants and tar.
Limita de detecţie:
≤ 10 µg/L
Material:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Metoda de măsurare:
Colorimetric measurement using DMG colour solution
Model:
EZ2727sc
Nr. flux probe :
Standard: 1
Optional accessory for 2, 4, or 8 sample streams
Parametru:
Nickel, total
Precizie:
Better than 2% full scale range for standard test solutions
Putere:
110 VAC / 60 Hz
Reagent Requirements:
Keep between 10 - 30 °C
Scurgere:
Atmospheric pressure, vented, min. Ø 32 mm
Temparatura ambiantă:
10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura probă:
10 - 30 °C
Validation:
Automatic; frequency freely programmable