EZ7900 Toxicity Analyser

EZ7900 Toxicity Analyser
Nr. produs: EZ7900.XXXXXXXX
Preț unitar Contact

Conţinutul cutiei

EZ7900 Toxicity Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and empty Containers for Reagents with Fittings

Specificaţii

Alarma: 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Calibrare: Automatic; standard at every analysis cycle
Calitate proba: Maximum particle size 500 µm, < 0.1 g/L
Most applications require the use of a dual EZ9110 + EZ9120 sampling/filtration system.
Certificări: CE compliant / ETL certified
Clasa de protectie: Analyser cabinet: IP44 / Panel PC: IP65
Conţinutul cutiei: EZ7900 Toxicity Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and empty Containers for Reagents with Fittings
Curatare automata: Da
Cuve: 10 - 15 minutes
Debitul probei: 100 - 300 mL/min
Demineralised water: N.A.
Digital outputs: Optional: Modbus (TCP/IP, RS485)
Dimensiuni (Înălţime x Lăţime x Adâncime): 690 mm x 465 mm x 330 mm
Domeniu de măsurare: 0 - 100 % inhibition
  0 - 100 % inhibition
Earth connection: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Esantion: presiune: By external overflow vessel
Garanţie: 24 luni
Greutate: 25 kg
Iesiri analogice: Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferente: N.A.
Limita de detecţie: N.A.
Material: Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Metoda de măsurare: Respirometry
Nr. flux probe : 1, 2, or 4
Parametru: Toxicity; respiration rate
Precizie: Better than 5% full scale range for standard test solutions
Putere: 100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Reagent Requirements: Keep between 10 - 30 °C
Scurgere: Atmospheric pressure, vented, min. Ø 32 mm
Temparatura ambiantă: 10 - 30 °C ±4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura probă: 10 - 30 °C
Validation: Automatic; frequency freely programmable